| 300mA & 500 mA |
| |
 |
| PICKS 3010 D, 300mA, 100/125Kv DIAGNOSTIC X-RAY GENERATOR. |
| |
Picks 3010 D X-Ray generator is highly suitable for all radiography, fluoroscopy, spot filming, image intensifier with TV system and other special examinations with manual and motorised Bucky Tables. |
|
 |
|
SALIENT FEATURES
|
- Two pulse single phase X-ray generator with maximum rating 300mA and 100/125Kv (24Kw)
- Full wave rectifications.
- Shorter exposure time 10ms.
- Digital display of mA, Kv, & mAs.
- Automatic computation of mAs digitally.
- Kv display shows F-Kv ON spot mode & R-Kv-Off spot mode.
- Overload protection to block the exposure and R-O-L & F-O-L glow.
- Rapid rotor anode braking circuit (to increase tube life).
- Continuous fluoroscopy by pressing F-RST push button.
- Double mA range compact meter with double lamp.
- LED lamps to indicate focii (different focus of the tube).
- Spot film on any technic is possible.
- High voltage silicon rectifier for low losses and consistent radiographic results.
|
TECHNICAL DATA
|
| Power Supply |
Nominal 415 V at 0.5Ω power line resistance 50/60 Hz. |
| Output |
300mA and 100/125Kvp [24Kw as per IS 7620-1986] |
| Fluoroscopy |
40Kv to 100Kv in 8 steps |
| Radiography |
40Kv to 115Kv in 8 steps.
0Kv to 10Kv in 6 steps.
On small focus 50, 100, 150mA.
On large focus 100, 200, 150mA. |
| Timer |
10ms to 5s in 23 steps. |
| X-Ray Tube |
Single Tube/Double Tube Compatible with BELDRA-1-125/20/40/Equivalent imported tubes with higher rating. |
|
|
| |
| PICKS 5012 D, 500mA, 125Kv DIAGNIOSTIC X-RAY GENERATOR |
| |
Picks 5012 D X-Ray generator is highly suitable for all radiography, fluoroscopy, spot filming, image intensifier with TV system and other special examinations with manual and motorised Bucky Tables. |
|
 |
|
SALIENT FEATURES
|
- Two pulse single phase X-ray generator with maximum rating 500mA and 125Kv (40Kw).
- Full wave rectifications.
- Shorter exposure time 10ms.
- Digital display of mA, Kv, & mAs.
- Automatic computation of mAs digitally.
- Kv display shows F-Kv ON spot mode & R-Kv-Off spot mode.
- Overload protection to block the exposure and R-O-L & F-O-L glow.
- Rapid rotor anode braking circuit (to increase tube life).
- Continuous fluoroscopy by pressing F-RST push button.
- Double mA range compact meter with double lamp.
- LED lamps to indicate focii (different focus of the tube).
- Solid state X-ray contacting system in lieu of elctro-magnetic contacting system.
- Spot film on any technic is possible.
- High voltage silicon rectifier for low losses and consistent radiographic results.
|
TECHNICAL DATA
|
| Power Supply |
Nominal 415 V at 0.3Ω power line resistance 50/60 Hz. |
| Output |
500mA and 125/125Kvp [40Kw as per IS 7620-1986] |
| Fluoroscopy |
40Kv to 100Kv in 8 steps |
| Radiography |
40Kv to 115Kv in 8 steps.
0Kv to 10Kv in 6 steps.
On small focus 50, 100, 150mA.
On large focus 100, 200, 300,500mA. |
| Timer |
10ms to 5s in 23 steps. |
| X-Ray Tube |
Single Tube/Double Tube Compatible with BEL DRA-1-125/20/40/Equivalent imported tubes with higher rating. |
|
 |
 |
X-Ray with motorised table |
Retrofit |
|
|
|